Model
|
2307
|
|
Ref.
|
30105
|
|
Class and Type
|
Portable instrument with interhal power supply
|
|
Applied Part
|
Type CF, protected against defibrillator discharges
|
|
Possible measurements
|
Sensing values, pacing threshold.
Impedance of the pacing circuit at 5V
|
|
Sensitivity
|
0,2 ÷ 50mV peak-peak;
Tolerance: the highest between ± 0,1mV and ±15%
|
|
Spontaneous frequency
|
15 ÷ 230 bpm;Tolerance: the highest between ± 1bpm and ± 5%
|
|
Pulse ampitude
|
0,1 ÷ 15V adjustable with 0,1V steps;
Tolerance: the highest between ± 0,1mV and ±15%
|
|
Pulse duration
|
0,25 ÷ 2,00 ms adjustable with 0,05 ms steps;
Tolerance: ± 10 %
|
|
Stimulation frequency
|
30 ÷ 180 ppm adjustable with 1 ppm steps;
Tolerance: ± 5 %
|
|
Stimulation method
|
Asynchronous and on demand
|
|
5V impedance (0,5 ms)
|
100 ÷ 3000 Ω ; Tolerance: ± 15 %
|
|
Emergency functions
|
Ventricular stimulation: Stimulation parameters preset at 60 ppm, 15V, 1 ms
High frequency a trial stimulation: 180 ÷ 1000 ppm
|
|
Display
|
LCD graphics,dimensions 60 x 42 mm, with possibility of rear illumination
|
|
Recommended threshold cable ( sterile)
|
Ref. F7830 S Sterile disposable threshold cable
|
|
Optional accessories
|
Ref. 30106 Infrared Printer (examination report on printout in heat-sensitive paper)
|
|
Power supply
|
Two 9V alkaline batteries ( mod. 6LR61)
|
|
Flat batteri signal
|
indication on the display
|
|
Autonomy
|
8 hours @ 20°C
|